<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>22003</ErrorNumber>
<ErrorFamily>Autoloader 2</ErrorFamily>
<ErrorMessage>
ALE2: No Plate sensed on table, nothing to load!
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>ThermoFlex</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
Check the table plate sensor to make sure it is plugged in and functional.
Use the 'table' command to query the plate sensor status, tape or your hand
will simulate a plate.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Make sure plate edge is straight and covers the table plate sensor when
loading.
</text></DescriptionText>
</OperatorAction>
<Cause>
In a load situation, the first push of the START button is to activate the vacuum
on the drum.  The firmware looks to see if there is a plate present or not before
it actually activates the vacuum.  In this case it failed to see a plate on the
table and it's complaining.
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../commands/table.xml</LinkAddress><LinkTitle>Command: table</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:50 2008 PST
</Updated>
<Links>
</Links>
</Error>